Home Home Patient Monitoring Multi Parameter Monitors

Riester Vital Signs Monitor RVS-100 with NIBP and SPO2

RI.1960-RRXXA

EACH
NZ $3,795.00
 Overseas Supplier Stock - Can take 3-4 weeks to arrive

Overview

Riester Vital Signs Monitor RVS-100

Advanced vital signs monitor for pulse oximetry, blood pressure and temperature measurements.

Riesters RVS-100 is an advanced vital signs monitor which offers pulse oximetry, blood pressure and temperature measurements. The device has Monitor, Spot-Check and Triage profile options and a high-performance touchscreen.

The RVS - 100 can communicate through a wired or wireless connection to the hospitals EMR, according to the HL7 standard. The highly flexible, modular design offers many configurations and settings to suit the needs of hospitals, clinics, and private practices, as well as long-term care settings.

Product Details

  • 3 measurement modes (Monitor, Spot-check, Triage)
  • High-sensitivity 8 TFT colour touch screen with self-explanatory user interface and integrated help- function
  • Displays SYS, DIA, MAP, Pulse, Temp, Plethysmogram
  • Clinical grade NIBP* including averaging mode
  • Programmable NIBP averaging mode (2-5 readings)
  • Riester SpO2 as Standard. (Can be upgraded to Nellcor Oximax or Masimo(POA))
  • Predictive Thermometer (Filac 3000) RI.1960-RRBXE only Model
  • Connectivity with the EMR - follows the HL7 standard
  • LAN Connectivity as standard
  • Programmable alarms (audible, visible) with 3 priority levels
  • Supports nurse call RJ11 port
  • Available as wall, desk and floor model
  • Can be expanded to wall diagnostic station
  • Internal memory for 5,000 measurements
  • Lithium-ion battery with up to 17 hours ON time

Options

  • Thermal printer (internal)
  • Barcode reader
  • Mobile stand RI.13317 including basket for storage (cuffs, sensors, cables)
  • WiFi


Accessories supplied withmonitor

  • NIBP extension tube
  • 2 cuffs (adult & adult large incl. connectors)
  • SpO2 sensor adult
  • SpO2 extension cable

Extra Accessories on RI.1960-RRBXE

  • Oral temperature probe
  • Probe cover

Technical Data

Monitor

Power Supply: 100 240 V AC, 50/60 Hz

Battery:

  • Lithium-Ion battery, rechargeable, 10.8 V DC, 6,600 mAh,
  • Charging time to a 100%: 6 hours,
  • Operating time approx. 11 hours (new and fully charged battery at 25C ambient temperature with connected SpO2, Temp, and NIBP on AUTO mode (15 minutes interval))
  • Low battery indicator
  • Charging battery indicator

Dimensions Monitor: L x H x W: 303 x 230 x 110 mm (11.93 x 9.06 x 4.33 in)

Weight: 3.60 kg (7.94 lb) including temp module, printer

Dimensions Basket: L x H x W: 270 x 225 x 200 mm (10.63 x 8.86 x 7.88 in)

Alarm:

  • Visible LEDs (Yellow, Red, Blue, Green, Orange)
  • Audible Speaker
  • Gives audible alarm, QRS tone
  • Supports Pitch Tone and multi-level volume
  • Alarm tones meet the requirement of IEC 60601-1-8.
  • Alarm Pressure: 45 dB to 85 dB test distance is 1 meter from the tone.

Ports:

  • 1 x AC power inlet
  • 1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
  • 4 x USB ports
  • 1 x Equipotential grounding point
  • 1 x RJ11 connector for nurse call
  • DC out port 15V/1.2 A

Clinical-grade NIBP (non-invasive blood pressure)

Measuring method: automatic oscillometric

Measurement range:
Adult SYS 30 270 mmHg, DIA 10 220 mmHg, MAP 20 235 mmHg
Pediatric SYS 30 235 mmHg, DIA 10 220 mmHg, MAP 20 225 mmHg
Neonate SYS 30-135 mmHg, DIA 10-110 mmHg, MAP 20-125 mmHg

Cuff pressure range: 0 280 mmHg

Resolution: 1 mmHg

Pressure accuracy:

  • Static 3 mmHg
  • Clinic 5 mmHg, standard deviation 8 mmHg
  • Pulse rate range: 40 240 bpm
  • Pulse rate accuracy: 1 bpm

Initial inflation pressure:

  • Adult 160 mmHg, Pediatric 130 mmHg, Neonate 75 mmHg
  • Double overpressure protection (hardware and software)
  • Adult 297 3 mmHg, Pediatric 252 3 mmHg , Neonate 147 3 mmHg

Alarm range:

  • SYS 0 300 mmHg
  • DIA 0 300 mmHg
  • MAP 0 300 mmHg

Riester SpO2

  • Measurement range: 0 100%
  • Resolution: 1%
  • Accuracy: 2% (70 to 100%)
  • 3% (40% to 69%)
  • Unspecified (0% to 39%)
  • Alarm range: 0 100%
  • Average time: 4 sec, 8 sec, 16 sec

Pulse rate

  • Measurement range: 20 250 bpm
  • Resolution: 1 bpm
  • Accuracy: 1% or 1 bpm whichever is greater

Predictive thermometer for RI.1960-RRBXE

  • Oral ( Quick Mode )
  • non-fever temps: 3-5 seconds
  • fever temps: 8-10 seconds
  • Oral ( Standard Mode ): 6-10 seconds
  • Axillary Mode: 8-12 seconds
  • Rectal Mode: 10-14 seconds
  • Direct Mode (all sites): 60-120 seconds
  • Measurement accuracy: 35.5C - 42C (95.9F 107.6F)
  • Direct Mode (all sites): 0.1C (0.2F)
  • Standard Prediction Mode (all sites): 0.1C (0.2F)
  • Quick Prediction Mode (oral only): 0.3C (0.5F)

Questions?