Orders Update

Please be advised that there is a slight delay on orders being dispatched. We are committed to getting these orders out to you as swiftly as possible and appreciate your patience and understanding during this period.
Home Home Intravenous & Administration IV Pump

Alaris neXus CC Syringe Pump

CCNEXUS1-S

EACH
NZ $4,284.90
 Ordered In as Required - Usually Ships Within 2-3 Days

Overview

Alaris neXus CC Syringe Pump

The Next-Gen BD syringe pump with flexible In-Line-Pressure Monitoring and Wi-Fi connectivity to support proactive vigilance in I.V. medication delivery.

The BD Alaris neXus CC Syringe Pump offers SMART technology to recognise if a pressure sensing disc is in place. This allows clinicians the flexibility to use either a conventional I.V. extension set or the dedicated In-Line-Pressure Monitoring set if situation requires.

Features and Benefits

  • Activation of In-Line Pressure Monitoring on demand, supports early intervention to help prevent potential complications related to occlusions
  • Wi-Fi connectivity enables automatic transmitting of infusion data to the hospital Clinical Information Systems.
  • Flexible END RATE for Volume To Be Infused (VTBI) with options to Stop, Keep Vein Open (KVO) or Continue at the set rate, supporting workflow efficiency by minimising disruptions to critical infusions (e.g. Inotropes).
  • Ability to program a loading dose and additional new dose units (kcal, mEq)
  • Support infusion precision with BACK OFF - an automatic feature which releases the pressure built up in the infusion system following an occlusion and minimises the post occlusion bolus.
  • FAST START a feature to automatically reduce the mechanical slack between the plunger mechanism and syringe at the start of an infusion.
  • Extensive drug library capacity to help standardise protocols hospital wide, 3,000 drug and fluid setups, spread over 30 available profiles, no additional constraint on the number of setups per profile.
  • BD Alaris MRI capsule is available to support continuity of IV therapy during imaging situations
  • BD Alaris Gateway Workstation creates an organised workspace and connects pumps to the hospitals Clinical Information Systems.
  • More resilient casing material designed to improve durability to a variety of cleaning agents.

Technical

  • Infusion rate range
  • 5 mL syringe 0.1 to 150 mL/h; 10 mL syringe 0.1 to 300 mL/h; 20 mL syringe 0.1
  • to 600 mL/h; 30 mL syringe 0.1 to 900 mL/h; 50 mL syringe 0.1 to 1200 mL/h;
  • increments starting at 0.01 mL/h.
  • Volume to be infused (VTBI)
  • 0.1 mL to 1000 mL, 1 minute to 24 hours.
  • Volume infused range
  • 0.1 to 9,990 mL.
  • Bolus rate
  • 5 mL syringe 10 to 150 mL/h; 10 mL syringe 10 to 300 mL/h; 20 mL syringe
  • 10 to 600 mL/h; 30 mL syringe 10 to 900 mL/h; 50 mL syringe 10 to 1200 mL/h.
  • Purge specifications (syringe size specific)
  • 100 to 500 mL/h; volume range 0.5 - 5.0 mL.
  • Keep vein open (KVO) infusion rate
  • 0.1 to 2.5 mL/h.
  • Flow rate accuracy (mL/h mode)
  • 2% for rates 1.0 mL/h in accordance with IEC 60601-2-24.#
  • Pumping pressure limits (user selectable)
  • In-line pressure measurement with pressure disc; variable pressure;
  • 1 to 1000 mmHg in 1 mmHg steps; auto pressure; 12 hour pressure trend graph.
  • Dimensions and weight
  • 335 mm (w) x 121 mm (h) x 200 mm (d).
  • 2.4 kg including carry handle, internal battery and integrated pole clamp.

Questions?

In order to assist us in reducing spam, please type the characters you see: